A moving-table Mössbauer spectrometer

J. H. Broadhurst, G. H. Guest, D. A. O'Connor, E. C. Shakespeare, H. R. Shaylor

Research output: Contribution to journalArticlepeer-review

8 Scopus citations

Abstract

A motion device for Mössbauer experiments is described consisting of a large moving stage supported on air bearings and driven by a servo-controlled hydraulic piston. The displacement of the table is detected using optical gratings. Velocity stability and reproducibility is within 0.1 % over a range 0.01 to 25 mm s-1.

Original languageEnglish (US)
Pages (from-to)275-279
Number of pages5
JournalNuclear Instruments and Methods
Volume73
Issue number3
DOIs
StatePublished - Sep 15 1969

Bibliographical note

Funding Information:
This work was supported by grants from the R.W. Paul Scientific Instrument Fund of the Royal Soci- ety of London and the Science Research Council. The authors wish to thank Prof. P.B. Moon for his advice and encouragement.