An unusual cause of asystole during a routine pacemaker generator change

Research output: Contribution to journalArticlepeer-review

Abstract

A 74-year-old pacemaker-dependent male patient, who underwent a pacemaker generator change due to elective replacement indicator, had a 4-second pause during interrogation of the new pacemaker generator out of the device pocket due to a specific feature that was programmed on called "enhanced transtelephonic monitoring." (PACE 2009; 32:949-951)

Original languageEnglish (US)
Pages (from-to)949-951
Number of pages3
JournalPACE - Pacing and Clinical Electrophysiology
Volume32
Issue number7
DOIs
StatePublished - Jul 2009

Keywords

  • Pacing
  • Transtelephonic monitoring

Fingerprint

Dive into the research topics of 'An unusual cause of asystole during a routine pacemaker generator change'. Together they form a unique fingerprint.

Cite this