Calibration and numerical simulation of Nanoparticle Surface Area Monitor (TSI Model 3550 NSAM)

W. G. Shin, D. Y.H. Pui, H. Fissan, S. Neumann, A. Trampe

Research output: Contribution to journalArticlepeer-review

98 Scopus citations

Abstract

TSI Nanoparticle Surface Area Monitor (NSAM) Model 3550 has been developed to measure the nanoparticle surface area deposited in different regions of the human lung. It makes use of an adjustable ion trap voltage to match the total surface area of particles, which are below 100 nm, deposited in tracheobronchial (TB) or alveolar (A) regions of the human lung. In this paper, calibration factors of NSAM were experimentally determined for particles of different materials. Tests were performed using monodisperse (Ag agglomerates and NaCl, 7-100 nm) and polydisperse particles (Ag agglomerates, number count mean diameter below 50 nm). Experimental data show that the currents in NSAM have a linear relation with a function of the total deposited nanoparticle surface area for the different compartments of the lung. No significant dependency of the calibration factors on particle materials and morphology was observed. Monodisperse nanoparticles in the size range where the response function is in the desirable range can be used for calibration. Calibration factors of monodisperse and polydisperse Ag particle agglomerates are in good agreement with each other, which indicates that polydisperse nanoparticles can be used to determine calibration factors. Using a CFD computer code (Fluent) numerical simulations of fluid flow and particle trajectories inside NSAM were performed to estimate response function of NSAM for different ion trap voltages. The numerical simulation results agreed well with experimental results.

Original languageEnglish (US)
Pages (from-to)61-69
Number of pages9
JournalJournal of Nanoparticle Research
Volume9
Issue number1
DOIs
StatePublished - Jan 2007

Keywords

  • Alveolar
  • Deposition in compartments of human lung
  • Instrumentation
  • Nanoparticle surface area
  • Occupational health
  • Tracheobronchial

Fingerprint

Dive into the research topics of 'Calibration and numerical simulation of Nanoparticle Surface Area Monitor (TSI Model 3550 NSAM)'. Together they form a unique fingerprint.

Cite this