Effectiveness of Phototherapy Units in Cameroon

Research output: Contribution to journalArticlepeer-review

9 Scopus citations

Abstract

Neonatal hyperbilirubinemia continues to be a leading cause of morbidity and mortality in resource-limited countries. The aim of this study was to measure the effectiveness of existing phototherapy units at a local hospital in Cameroon using an irradiance meter. Phototherapy units (n 1/4 4) in one newborn nursery in Cameroon were evaluated. The average irradiance of the functioning units was 2.87 mW/cm2/nm, which is substantially below the recommended range of 10-30 mW/cm2/nm. With simple improvements, one new prototype unit was developed. Its irradiance was 23.3 mW/cm2/nm. We concluded that irradiance of phototherapy units should be measured, as many local nurseries worldwide may not be delivering effective treatment. Simple and costeffective changes to phototherapy units can make a substantial improvement in irradiance.

Original languageEnglish (US)
Article numberfmt110
Pages (from-to)264-266
Number of pages3
JournalJournal of Tropical Pediatrics
Volume60
Issue number3
DOIs
StatePublished - May 2014

Keywords

  • Newborn
  • Phototherapy
  • jaundice

Fingerprint Dive into the research topics of 'Effectiveness of Phototherapy Units in Cameroon'. Together they form a unique fingerprint.