High-speed simultaneous determination of nine antiepileptic drugs using liquid chromatography-mass spectrometry

Murali Subramanian, Angela K. Birnbaum, Rory P. Remmel

Research output: Contribution to journalArticlepeer-review

102 Scopus citations

Abstract

Therapeutic drug monitoring of antiepileptic drugs (AEDs) is important and widely practiced. However, simultaneous AED assays usually concentrate only on old or new AEDs. A new simultaneous assay was developed to monitor both older and newer AEDs in the same sample. This assay measures zonisamide (ZNS), lamotrigine (LTG), topiramate (TPM), phenobarbital (PB), phenytoin (PHT), carbamazepine (CBZ), carbamazepine-10,11-diol (CBZ-Diol), 10- hydroxycarbamazepine (MHD), and carbamazepine-10,11-epoxide (CBZ-E). Sample pretreatment consisted of a single solid-phase extraction (SPE) for all AEDs in a 100-μL plasma sample. HPLC separation was achieved on a Shimdazu Shimpack XR-ODS (4.6 id × 50 mm, 2.2-μm) column with a gradient mobile phase of acetate buffer, methanol, acetonitrile, and tetrahydrofuran. Four internal standards were used. Detection was achieved by atmospheric pressure chemical ionization mass spectrometry (APCI-MS) in selected-ion monitoring (SIM) mode with constant polarity switching. High recovery (88%-96%) was obtained for all compounds by SPE. Linearity was observed throughout an 80-fold concentration range, with correlation coefficient (r 2) values higher than 0.99 for all AEDs. For the standards, the accuracy ranged from 89.3% to 111.8%. The within-run coefficient of variation (CV w) value was ≤9.7%, the between-run coefficient of variation (CV b) was ≤16.2%, and the total variability (CV t) was ≤16.8%. For the quality controls (QCs), accuracy ranged from 89.3% to 108.8%, CV w was ≤9.6%, CV b was ≤14.1%, and CV t was ≤15.1%. The correlation r 2 values for comparison of this assay with existing validated assays in our laboratory (GC-MS or LC-MS) were 0.95, 0.91, 0.87, 0.95, and 0.95 for PHT, LTG, CBZ, CBZ-E, and CBZ-Diol, respectively.

Original languageEnglish (US)
Pages (from-to)347-356
Number of pages10
JournalTherapeutic drug monitoring
Volume30
Issue number3
DOIs
StatePublished - Jun 2008

Keywords

  • Antiepileptic drugs (AEDs)
  • Carbamazepine
  • LC-MS
  • Lamotrigine
  • Simultaneous assay
  • Solid-phase extraction
  • Zonisamide

Fingerprint

Dive into the research topics of 'High-speed simultaneous determination of nine antiepileptic drugs using liquid chromatography-mass spectrometry'. Together they form a unique fingerprint.

Cite this