In vitro comparison of nitrofurazone- and silver alloy-coated foley catheters for contact-dependent and diffusible inhibition of urinary tract infection-associated microorganisms

James R. Johnson, Brian Johnston, Michael A. Kuskowski

Research output: Contribution to journalArticlepeer-review

44 Scopus citations

Abstract

Two marketed antimicrobial-coated Foley catheters were compared for in vitro diffusible and contact-dependent inhibition of 11 urinary tract infection-associated microorganisms in an adherence-biofilm assay. Nitrofurazone-coated catheters significantly outperformed silver alloy-coated catheters for inhibitory activity, according to both inoculum broth and catheter sonicate counts, whether compared directly or against the corresponding control catheters. Although inhibition waned with catheter preincubation in saline, some organisms were inhibited even after a 48-h catheter preincubation, especially by the nitrofurazone-coated catheter.

Original languageEnglish (US)
Pages (from-to)4969-4972
Number of pages4
JournalAntimicrobial agents and chemotherapy
Volume56
Issue number9
DOIs
StatePublished - Sep 2012

Fingerprint

Dive into the research topics of 'In vitro comparison of nitrofurazone- and silver alloy-coated foley catheters for contact-dependent and diffusible inhibition of urinary tract infection-associated microorganisms'. Together they form a unique fingerprint.

Cite this