Long-term stability of circumferential silicone sponge scleral buckling exoplants

David M. Brown, Robert M. Beardsley, Richard H. Fish, Tien P. Wong, Rosa Y. Kim

Research output: Contribution to journalReview articlepeer-review

18 Scopus citations

Abstract

PURPOSE: The historical extrusion/explantation rate of silicone sponge scleral buckle exoplants is 3.5 to 24.4%. This contrasts with the published 0.6 to 1.2% explantation rate of solid silicone elements. Previously reported silicone sponge exoplants studies were noncircumferential (quadrantic or multiple). This study was undertaken to assess the long-term stability of 360° circumferentially placed 3 × 5 mm silicone sponge exoplants. METHODS: Interventional case series of 840 consecutive circumferentially placed 3 × 5 mm silicone sponge exoplants. A retrospective review of operative reports and patient charts was performed. RESULTS: A total of 552 patients had documented follow-up over 1 year. Median follow-up was 32.8 months. Six patients underwent removal of the scleral buckling element (two for diplopia, one acute postoperative infection, and three for extrusion). Median time to removal was 7.3 months. Extenuating circumstances contributed to two of the three extrusions. Of the entire series, fewer than 1% of patients required removal of the silicone sponge implant. Excluding patients with less than 1 year of documented follow-up, the long-term infection/extrusion rate was 0.7%. CONCLUSIONS: Circumferentially placed silicone scleral buckling sponge elements are very stable and are well tolerated. The explantation/extrusion rate is comparable to published solid silicone element series.

Original languageEnglish (US)
Pages (from-to)645-649
Number of pages5
JournalRetina
Volume26
Issue number6
DOIs
StatePublished - Jul 2006

Keywords

  • Retinal detachment repair complications
  • Rhegmatogenous retinal detachment repair
  • Scleral buckle
  • Scleral buckle extrusion/infection rates
  • Silicone sponge exoplant

Fingerprint

Dive into the research topics of 'Long-term stability of circumferential silicone sponge scleral buckling exoplants'. Together they form a unique fingerprint.

Cite this