Megavoltage cone beam computed tomography dose and the necessity of reoptimization for imaging dose-integrated intensity-modulated radiotherapy for prostate cancer

Yuichi Akino, Masahiko Koizumi, Iori Sumida, Yutaka Takahashi, Toshiyuki Ogata, Seiichi Ota, Fumiaki Isohashi, Koji Konishi, Yasuo Yoshioka

Research output: Contribution to journalArticlepeer-review

11 Scopus citations

Abstract

Purpose: Megavoltage cone beam computed tomography (MV-CBCT) dose can be integrated with the patient's prescription. Here, we investigated the effects of imaging dose and the necessity for additional optimization when using intensity-modulated radiotherapy (IMRT) to treat prostate cancer. Methods and Materials: An arc beam mimicking MV-CBCT was generated using XiO (version 4.50; Elekta, Stockholm, Sweden). The monitor units (MU) for dose calculation were determined by conforming the calculated dose to the dose measured using an ionization chamber. IMRT treatment plans of 22 patients with prostate cancer were retrospectively analyzed. Arc beams of 3, 5, 8, and 15 MU were added to the IMRT plans, and the dose covering 95% of the planning target volume (PTV) was normalized to the prescribed dose with (reoptimization) or without optimization (compensation). Results: PTV homogeneity and conformality changed negligibly with MV-CBCT integration. For critical organs, an imaging dose-dependent increase was observed for the mean rectal/bladder dose (D mean), and reoptimization effectively suppressed the D mean elevations. The bladder generalized equivalent uniform dose (gEUD) increased with imaging dose, and reoptimization suppressed the gEUD elevation when 5- to 15-MU CBCT were added, although rectal gEUD changed negligibly with any imaging dose. Whereas the dose elevation from the simple addition of the imaging dose uniformly increased rectal and bladder dose, the rectal D mean increase of compensation plans was due mainly to low-dose volumes. In contrast, bladder high-dose volumes were increased by integrating the CBCT dose, and reoptimization reduced them when 5- to 15-MU CBCT were added. Conclusion: Reoptimization is clearly beneficial for reducing dose to critical organs, elevated by addition of high-MU CBCT, especially for the bladder. For low-MU CBCT aimed at bony structure visualization, compensation is sufficient.

Original languageEnglish (US)
Pages (from-to)1715-1722
Number of pages8
JournalInternational Journal of Radiation Oncology Biology Physics
Volume82
Issue number5
DOIs
StatePublished - Apr 1 2012

Bibliographical note

Copyright:
Copyright 2012 Elsevier B.V., All rights reserved.

Keywords

  • Dose compensation
  • Megavolt cone beam computed tomography
  • Prostate IMRT
  • Reoptimization

Fingerprint

Dive into the research topics of 'Megavoltage cone beam computed tomography dose and the necessity of reoptimization for imaging dose-integrated intensity-modulated radiotherapy for prostate cancer'. Together they form a unique fingerprint.

Cite this