Abstract
Acyclovir, ganciclovir and (R)-9-[4-hydroxy-2-(hydroxymethyl)butyl]guanine are active in vitro against the Epstein-Barr virus (EBV) but their in vivo anti-EBV activity is not well understood. We developed a novel, sensitive high-performance liquid chromatography assay with ultraviolet detection for measuring acyclovir, ganciclovir and (R)-9-[4- hydroxy-2-(hydroxymethyl)butyl]guanine in human plasma to identify quantitative relationships between in vitro anti-EBV activity and therapeutic response. Characteristics of the assay include a low plasma volume (200 μL), perchloric acid protein precipitation, use of penciclovir as the internal standard, run times less than 8 min and a 50 ng/mL lower limit of quantification. The within- and between-assay variability is 0.7-4.8 and 1.0-7.9%, respectively. Accuracy for all three drugs ranges from 89.5 to 106.4% for four quality controls (50, 100, 1000 and 10,000 ng/mL). This assay supports pharmacokinetic and pharmacodynamic studies of candidate anti-EBV drugs in children and adults with EBV infections.
Original language | English (US) |
---|---|
Pages (from-to) | 822-827 |
Number of pages | 6 |
Journal | Biomedical Chromatography |
Volume | 23 |
Issue number | 8 |
DOIs | |
State | Published - 2009 |
Keywords
- Acyclovir
- Chromatography
- Epstein-Barr virus
- Ganciclovir
- H2G