Threshold Measurement for Patients with Tinnitus: Pulsed or Continuous Tones

Suzette M. Mineau, Robert S. Schlauch

Research output: Contribution to journalArticlepeer-review

14 Scopus citations

Abstract

This study compared the effectiveness of pulsed tones (200 ms on/200 ms off) and continuous tones (1 to 2 s) for measuring thresholds in patients with tinnitus. Sixty veterans with tonal tinnitus and a history consistent with noise-induced hearing loss participated. Half were assigned to a pulsed-tone group; half were assigned to a continuous-tone group. Air-conduction thresholds were measured twice for each patient in a single session for octave intervals between 250 Hz and 4000 Hz. Results indicated no significant difference between groups for test-retest differences, but overall more presentations were required to measure thresholds with continuous tones than with pulsed tones at 4000 Hz. Moreover, the continuous-tone group had, on average, more false-positive responses than the pulsed-tone group. This study supports the clinical belief that pulsed tones are preferred over continuous tones for measuring thresholds for persons with tinnitus.

Original languageEnglish (US)
Pages (from-to)52-56
Number of pages5
JournalAmerican Journal of Audiology
Volume6
Issue number1
DOIs
StatePublished - Mar 1997

Keywords

  • Hearing loss
  • Threshold
  • Tinnitus

Fingerprint

Dive into the research topics of 'Threshold Measurement for Patients with Tinnitus: Pulsed or Continuous Tones'. Together they form a unique fingerprint.

Cite this