Impedance cardiography: What is the source of the signal?

Research output: Contribution to journalConference articlepeer-review

32 Scopus citations

Abstract

Impedance cardiography continues to be investigated for various applications. Instruments for its use are available commercially. Almost all of the recent presentations and articles along with commercial advertisements have assumed that aortic volume pulsation is the source of the signal. A review of the literature will reveal that there is no clear evidence for this assumption. Starting with the first paper on impedance cardiography in 1964, which assumed the lung was the source of the signal, the presentation will review many studies in the 60's, 70's and 80's, which suggest the aorta and other vessels as well as atria and again the lung as possible sources. Current studies based on high resolution thoracic models will be presented that show the aorta as contributing only approximately 1% of the total impedance measurement, making it an unlikely candidate for the major contributor to the signal. Combining the results of past studies along with recent work based on models, suggest other vessels and regions as possible sources.

Original languageEnglish (US)
Article number012118
JournalJournal of Physics: Conference Series
Volume224
Issue number1
DOIs
StatePublished - 2010
Event14th International Conference on Electrical Bioimpedance, Held in Conjunction with the 11th Conference on Biomedical Applications of EIT, ICEBI and EIT 2010 - Gainesville, FL, United States
Duration: Apr 4 2010Apr 8 2010

Fingerprint

Dive into the research topics of 'Impedance cardiography: What is the source of the signal?'. Together they form a unique fingerprint.