Kinetic spectrophotometric determination of sulfide using whole kinetic curve and a fixed time method

J. Ghasemi, D. E. Mohammadi

Research output: Contribution to journalArticlepeer-review

35 Scopus citations

Abstract

The univariate and multivariate calibration method was applied for the determination of sulfide based on the addition reaction of sulfide with new fuchsin at pH 8 and 25°C. The decrease in absorbance of new fuchsin at 540 nm over a fixed time is proportional to the concentration of sulfide over the range of 0.05-2.5 μg ml-1. The experimental calibration matrix for partial least squares calibration (PLS) method was designed with 20 samples. The cross-validation method was used for selecting the number of factors. The root mean square difference (RMSD) was 0.0843. The effects of various cations and anions on sulfide determination have been investigated.

Original languageEnglish (US)
Pages (from-to)1-8
Number of pages8
JournalMicrochemical Journal
Volume71
Issue number1
DOIs
StatePublished - 2002

Keywords

  • Determination
  • Fixed time method
  • New fuchsin
  • Non-catalytic
  • Partial least squares
  • Sulfide

Fingerprint