Lesion formation and visualization using dual-mode ultrasound phased arrays

Research output: Contribution to journalConference articlepeer-review

14 Scopus citations

Abstract

A 1 MHz spherical-section 64-element linear piezo-composite phased array and a supporting driving and data acquisition system were recently tested for use as a dual-mode HIFU applicator system. In therapeutic mode, the array was shown to be capable of producing focal intensities in excess of 3500 W/cm2 at its geometric center (100 mm radius of curvature). Imaging tests of quality assurance tissue-mimicking phantoms as well as computer simulations confirmed that the array has an oval-shaped imaging field of view (IxFOV) with nearly 50 dB dynamic range extending from 70 to 120 mm in the axial direction and ± 20 mm in the transverse direction. This is larger than its therapeutic operating field (ThxOF), which is defined as the region around the geometric focus where the loss in intensity gain is within 1 dB from the intensity gain at the geometric center. We have tested the array imaging capability in the visualization of the formation of discrete and volumetric lesions in freshly excised porcine liver samples. Experimental results clearly show that the enhancement of the tissue echogenecity at the lesion location is more pronounced in the harmonic images when compared to images at the fundamental. Results also clearly show that the harmonic imaging mode produces more accurate mapping of the lesion size and shape as determined by histologic evaluation.

Original languageEnglish (US)
Pages (from-to)1351-1354
Number of pages4
JournalProceedings of the IEEE Ultrasonics Symposium
Volume2
StatePublished - 2001
Event2001 Ultrasonics Symposium - Atlanta, GA, United States
Duration: Oct 6 2001Oct 10 2001

Fingerprint

Dive into the research topics of 'Lesion formation and visualization using dual-mode ultrasound phased arrays'. Together they form a unique fingerprint.

Cite this