Monitoring and guidance of HIFU beams with dual-mode ultrasound arrays

John R. Ballard, Andrew J. Casper, Emad S. Ebbini

Research output: Contribution to journalArticlepeer-review

4 Scopus citations

Abstract

We present experimental results illustrating the unique advantages of dual-mode array (DMUA) systems in monitoring and guidance of high intensity focused ultrasound (HIFU) lesion formation. DMUAs offer a unique paradigm in image-guided surgery; one in which images obtained using the same therapeutic transducer provide feedback for: 1) refocusing the array in the presence of strongly scattering objects, e.g. the ribs, 2) temperature change at the intended location of the HIFU focus, and 3) changes in the echogenicity of the tissue in response to therapeutic HIFU. These forms of feedback have been demonstrated in vitro in preparation for the design and implementation of a real-time system for imaging and therapy with DMUAs. The results clearly demonstrate that DMUA image feedback is spatially accurate and provide sufficient spatial and contrast resolution for identification of high contrast objects like the ribs and significant blood vessels in the path of the HIFU beam.

Fingerprint

Dive into the research topics of 'Monitoring and guidance of HIFU beams with dual-mode ultrasound arrays'. Together they form a unique fingerprint.

Cite this