Technical Requirements for Rapid High‐Efficiency Therapies

Prakash Keshaviah, Doug Luehmann, Karen Ilstrup, Allan Collins

Research output: Contribution to journalArticlepeer-review

29 Scopus citations

Abstract

Abstract: Significant reductions in treatment time are possible with rapid high‐efficiency hemodialysis and hemodiafiltration, provided the therapies are implemented so as to maintain adequacy of solute and fluid removal without compromising patient comfort. The key technical elements necessary for such implementation include high blood flow rates, higher efficiency dialyzers/diafilters, ultrafiltration control systems, and bicarbonate as the buffer source. In addition, hemodiafiltration requires schemes to ensure sterility and nonpyrogenicity of the infusion fluid and appropriate balancing of the rates of ultrafiltration and reinfusion. In general, rapid high‐efficiency therapies are technically more complex than standard therapy, and rapid hemodiafiltration appears to be more complex than rapid hemodialysis. The technology required for rapid hemodialysis currently exists, but it needs to be fine‐tuned and integrated for routine application.

Original languageEnglish (US)
Pages (from-to)189-194
Number of pages6
JournalArtificial Organs
Volume10
Issue number3
DOIs
StatePublished - Jun 1986
Externally publishedYes

Keywords

  • Bicarbonate buffer
  • High‐efficiency hemodiafiltration
  • High‐efficiency hemodialysis
  • Intradialytic symptoms
  • Reduced treatment time
  • Ultrafiltration control systems

Fingerprint

Dive into the research topics of 'Technical Requirements for Rapid High‐Efficiency Therapies'. Together they form a unique fingerprint.

Cite this